What to do with SI Error: "check path /usr/g/mrraw/HEADER_POOL

Symptoms:

After pressing 'Scan Setup' on the instascan, a dialog appears with the contents:

SI: Please check path
/usr/g/mrraw/HEADER_POOL

What Happens:

After certain procedures on the Signa (in particular, mrload) the access from the instascan to the Signa may become limited. This comes about when the contents of /etc/exports on the Signa revert to their factory configuration.

What to do:

Log on to the Signa and alter the contents of /etc/exports to include permission for instascan to mount /usr and /usr/g.

In particular, the altered file should read:

/            UC3T_OC0-tps       UC3T_OC0
/usr         UC3T_OC0-tps       UC3T_OC0 instascan
/usr/g       UC3T_OC0-tps       UC3T_OC0 instascan

After which, you must reboot the Signa (yes: reboot, not just restart the Signa programs...)