Setting up for Real Time Imaging
Establish a target directory for the Real-Time images
On the APD, edit the file:
/usr/export/home/anmr/config/.apdconfigMSCMOD
so that the line '
FILEPATH = xxx
' points to your target directory.
At the APD console, type "Faster" in a root window. This will bring down APD2 and restart it in Real-Time Mode.
At the Signa, set up an EPI sequence with the following parameters.
Sequence
ismemp or isgr
FOV
400
Matrix
128x128
Phase FOV
200
TE
Min. full or greater
When APD2 is up, set the scanner to
Research
mode
After the sequence is Downloaded on the Signa, Press
Scan Setup
.
Select the Experiment file,
fast.exp5x
Scroll down to the Read FOV, and set it to
200 mm.
Press
Proceed.
Carry on.
©
5/30/02
Mark S. Cohen, Ph.D
.