How to Restore the shim to standard 'Head' values

  1. From the instascan console, open a window and start the shimmer program.
  2. Select option 5 (more options)
  3. Select option 6 (read settings from a file)
  4. Enter in a file name. A good starting file is:
    /usr/export/home/anmr/stdhead
  5. After the file is loaded, select 7 (return to basic options)

A screen capture of this process appears below. User entries appear in boldface:

instascan# shimmer
e--------------------------------------------------------------------------e
| shimmer_r6 is a user-friendly interface for the suite of programs that |
| enable automated shimming on the Signa-Instascan systems. shimmer_r6 |
| was written by Tim Reese (me, the author). The programs it runs were |
| developed by Tim Davis (xds, WU), David Harrison (xgraph, UCB) and me, |
| with collaborative help from Robert Weisskoff. Please address questions, |
| change requests or bug reports to Tim Reese, reese@nmr.MGH.harvard.edu. |
e--------------------------------------------------------------------------e


----------- checking Resonance Research shim power supply----------------
opening serial port.
activating controller DAC.
controller power is currently on.
closing serial port.
WARNING --- shimmer:depthfirst_pwd:directory error at file .apdconfig:no such file or directory
--------------------------------------------------------------------------
X : -0.046144 ZX : 0.170049
Y : 0.140873 ZY : 1.197668
Z : 0.767724 XY : 0.077517
Z2 : -0.819727 X2Y2 : -0.257942
--------------------------------------------------------------------------

CURRENT PARAMETERS: Bay 2 Signa/Instascan, through 2nd order
columns, rows and slices: 128,64,9
voxel dimensions in mm: 3.2,3.2,10
Echo shift: 4.4 ms, Acquisition time: 32 ms
orientation: axial, I-S slice, R-L read, supine
current directory: /home/anmr
(no I/Q images found)

SHIM PROCESSING OPTIONS (TOP LEVEL MENU)
0) go process image data
1) <inactive>
2) <inactive>
3) start the manual shim controller
4) select a shim volume? no
5) more options
6) quit

(0 to 6, 0) :: 5

SHIM PARAMETER OPTIONS (SECOND LEVEL MENU)
0) pick shims to evaluate : Z0,Z1,X1,Y1,Z2,ZX,ZY,XY,X2Y2
1) enable cardiac mode? no (in development)
2) display correlations? no (in development)
3) change magnitude threshold from 10%
4) enable DUMP in shimfit? no
5) save shim settings to a file.
6) read shim settings from a file.
7) return to basic options

(0 to 7, 7) :: 6
Enter the file name to restore from (must be the full path!)
default: /tmp/shim_settings4029 :: /usr/export/home/anmr/stdhead
opening serial port.
closing serial port.
shimcontrol_r2: shim settings restored from file /usr/export/home/anmr/stdhead

SHIM PARAMETER OPTIONS (SECOND LEVEL MENU)
0) pick shims to evaluate : Z0,Z1,X1,Y1,Z2,ZX,ZY,XY,X2Y2
1) enable cardiac mode? no (in development)
2) display correlations? no (in development)
3) change magnitude threshold from 10%
4) enable DUMP in shimfit? no
5) save shim settings to a file.
6) read shim settings from a file.
7) return to basic options

(0 to 7, 7) :: 7